FUNDUS CAMERA
|
Type |
Non-mydriatic fundus camera |
Photography type |
Color |
Angle of View |
45° ± 5% |
Min. pupil size for Fundus |
3.3 mm |
Camera |
12.3 Megapixel |
Photography |
Fundus (Retina, Central, Disc, Peripheral with Manual fixation), Anterior |
Flash Adjustment, Gain, Exposure |
Auto, Manual |
Intensity levels |
High, Normal, Low |
OPTICAL COHERENCE TOMOGRAPHY
|
Technology |
|
Light Source |
SLED Wavelength 850 nm |
Bandwidth |
50 nm half bandwidth |
Scanning Speed |
130,000 measurements per second |
Axial Resolution |
2.8 μm digital, 5 μm in tissue |
Transverse Resolution |
12 μm, typical 18 μm |
Overall Scan Depth |
2.8 mm / ~6 mm in Full Range mode |
Min. Pupil Size for OCT |
2.4 mm |
Focus Adjustment Range |
-25 D to +25 D |
Scan Range |
Posterior 3 mm to 15 mm, Angio 3 mm to 15 mm, Anterior 3 mm to 18 mm |
Scan Types |
3D, Angio², Full Range Radial, Full Range B-scan, Radial (HD), B-scan (HD), Raster (HD), Raster 21 (HD), Cross (HD), TOPO ², Biometry AL ² |
Fundus Alignment |
IR, pSLO (Live Fundus Reconstruction) |
Alignment Method |
Fully automatic, Automatic, ManualFundus Tracking Real time active, iTracking |
Fundus Tracking |
ACCUtrack – active real time, iTracking |
Retina Analysis |
Retina thickness, Inner Retinal thickness, Outer Retinal thickness, RNFL+GCL+IPL thickness, GCL+IPL thickness, RNFL thickness, RPE deformation, MZ/EZ-RPE thickness |
Angiography OCT ² |
Vitreous, Retina, Choroid, Superficial Plexus, RPCP, Deep Plexus, Outer Retina, Choriocapilaries, Depth Coded, SVC, DVC, ICP, DCP, Custom, Enface, Quantification: FAZ, VFA, NFA, Vessel Area Density, Skeleton Area Density, Thickness map |
Glaucoma Analysis |
RNFL, ONH morphology, DDLS, OU and Hemisphere asymmetry, Ganglion analysis as RNFL+GCL+IP and GCL+IPL, Structure + Function¹ |
Angiography Mosaic |
Acquistion method: Auto, Manual Mosaic modes: 10×10, 10×6, 12×5, 7×7, Manual up to 12 images |
Biometry OCT ² |
AL, CCT, ACD, LT, P, WTW |
IOL Calculator ³ |
IOL Formulas: Hoffer Q, Holladay I, Haigis, Theoretical T, Regression II |
Corneal Topography Map ² |
Axial [Anterior, Posterior], Refractive Power [Kerato, Anterior, Posterior, Total], Net Map, Axial True Net, Equivalent Keratometer, Elevation [Anterior, Posterior], Height, KPI (Keratoconus Prediction Index) |
Anterior (no lens/adapter required) |
Anterior Chamber Radial, Anterior Chamber B-scan, Pachymetry, Epithelium map, Stroma map, Angle Assessment, AIOP, AOD 500/750, TISA 500/750, Angle to Angle view |
Connectivity |
DICOM Storage SCU, DICOM MWL SCU, CMDL, Networking |
Fixation Target |
OLED display (the target shape and position can be changed), External fixation arm |
Dimensions (LxWxH) / Weight |
479 mm × 367 mm × 493 mm / 30 kg |
Power Supply / Consumption |
100 V to 240 V, 50/60 Hz / 90 VA to 110 VA |